SYD-0653 Emulsified Asphalt Particles Ionic Charge Tester

Additional information

Maximum output load

30mA

Timing accuracy

0.1 seconds

Timing time

3min

Power supply voltage

220V

Dimensions

300*200*300mm

Weight

10KG

Power

200W

Ambient temperature

5~40℃

Ambient humidity

≦85﹪

Product details:

Usage

This instrument is used to determine the charge properties of ions of various emulsified asphalt particles, that is, the types of cations and anions. The particle ionic charge of emulsified asphalt refers to the charge properties of the particles dispersed in the emulsion to determine the type of emulsion. Positively charged (+) is cationic emulsified asphalt, and negatively charged (-) is anionic emulsified asphalt. By inserting an electrode into the asphalt emulsion and applying a DC 6V electric field, the instrument judges the polarity of the particle ion charge of the emulsified asphalt according to the movement and attachment of the ions.

 

Technical Parameters

  1. Power supply voltage: DC 6V
  2. Maximum output load: 30mA
  3. Timing accuracy: 0.1 seconds
  4. Timing time 3min
  5. Power supply voltage: 220V
  6. Dimensions: 300*200*300mm
  7. Weight: 10KG
  8. Power: 200W
  9. Ambient temperature: 5~40℃
  10. Ambient humidity: ≦85﹪

Ask For A Quick Quote

We will contact you within 1 working day, please pay attention to the email with the suffix “@gtj-test.com”